FreeScan UE - Calibration

This article was created using software version FreeScan UE_v1.2.2.4.

This article will give you information about the calibration of the scanner. It explains when does the scanner need to be calibrated and how to perform the calibration.  

When does the scanner need to be calibrated?

The scanner needs to be calibrated when:

  • using it for the first time
  • changing the PC
  • the scanner has problem to see the markers
  • you transport the scanner to another location
  • the temperature changed (for example the scanner was in the car overnight and in the morning, you take it to the office and want to start scanning)

Calibration workflow

The calibration consists of 5 steps. Before you start put the calibration board on a flat surface (table, floor). Ensure the calibration board is not damaged (no scratches, no missing markers, no dirt etc). When it's good, it should look like the pic bellow. 

A picture containing text, white, different

Description automatically generated

For the first calibration step hold the scanner parallel over the board, in the middle and go slowly up.

To start the calibration, you need to push the trigger one time. After the calibration is done the scanner will stop automatically. 

A picture containing watch

Description automatically generated

When starting the calibration, you should see that the LEDs around the cameras are shining, and the laser are shining as well. By moving the scanning left, right, front and back you will find the correct position that the bars on the left, on the right and the big circle in the middle are green. Like in the picture bellow.

A picture containing graphical user interface

Description automatically generated

Start the calibration from near to far until all bars on the right side are marked green. In the video bellow you see the workflow from the first step.

A picture containing indoor, person

Description automatically generated

Graphical user interface, application

Description automatically generated

After the first step is done the second calibration step need to be done. In these steps and for the rest the lasers are OFF and only the LEDs from the scanner are ON. For the second step, the scanner must be tilted forward like in the picture bellow.

Calendar

Description automatically generated

When you have tilted the scanner forward and find the correct position, please start from near to far until all bars on the right side are marked green. In the video bellow you can the the process.

A picture containing computer, indoor, computer

Description automatically generated

Graphical user interface, application

Description automatically generated

Now step 3 needs to be done. For this step the scanner must be tilted backward like in the picture bellow.

A picture containing diagram

Description automatically generated

When you have tilted the scanner backward start from near to far until all bars on the right side are marked green. In the video bellow you can the process.

A picture containing person, indoor, desk

Description automatically generated

Graphical user interface, application

Description automatically generated

Now step 4 needs to be done. For this step the scanner must be tilted to the right like in the picture bellow.

Diagram

Description automatically generated with low confidence

When you have tilted the scanner to the right, start from near to far until all bars on the right side are marked green. In the video bellow you can the process.

A picture containing person, indoor, worktable, desk

Description automatically generated

Graphical user interface

Description automatically generated with medium confidence

For the last step 5 the scanner must be tilted to the left like in the picture bellow.

A picture containing text, indoor, lighter

Description automatically generated

When you have tilted the scanner to the left, start from near to far until all bars on the right side are marked green. In the video bellow you can the process.

A picture containing person, indoor, worktable

Description automatically generated

Graphical user interface, application

Description automatically generated

After all these steps has been done the calibration is finishes and the software start to calculate the result. If everything was done correctly a new window should pop like bellow.

A screenshot of a computer

Description automatically generated with medium confidence

 

If the calibration fails, please try again. If it fails a second time please check with the following article, if your PC is set up correctly:

Set up your PC


When your PC is correctly set up but the calibration still fails, please contact us via metrology_support@shining3d.com.