FreeScan UE Pro - Calibration

This article was created using software version 1.2.1.5


This article will give you information about the calibration of the scanner. It explains when to calibrate and how to perform the calibration.


Watch a video tutorial!


1. When to calibrate


The scanner need to be calibrated when:

  • using it for the first time
  • changing the PC
  • the scanner has problem to see the markers
  • you transport the scanner to another location
  • the temperature changed (for example the scanner was in the car overnight and in the morning you take it to the office and want to start scanning)



2. Calibration workflow


The calibration consists of 6 steps. Before you start put the calibration board on a flat surface (table, floor). Ensure the calibration board is not damaged (no scratches, no missing markers, no dirt etc). When you put the board informt of you ensure that the left side (L) is left and right side (R) is right. When it's good, it should look like the pic bellow. 




For the first calibration step the 5 parelle lines will be calibrated. Hold the scanner parallel over the left side of the board. The distance is quite close. Hold it in the middle and go slowly up.

To start the calibration you need to push the trigger one time. After the calibration is done the scanner will stopp automatically. 




When starting the calibration you should see that the LEDs around the cameras are shining and in the first and second step the laser are shining as well. By moving the scanning left, right, front and back you will find the correct position that the bars on the left, on the right and the big circle in the middle are green. Like in the picture bellow.



Start the calibration from near to far until all bars on the right side are marked green. In the video bellow you see the workflow from the first stepp.




After the first step is done the 2 calibration step need to be done. Now all crossing line and single laser line are to ON as well as the LEDs. Hold the scanner in the middle of the board and move from near to far. Bellow you see the workflow .




After the second step is done the 3 calibration step need to be done. In this steps and also for the rest the lasers are OFF and only the LEDs from the scanner are ON. For the second step, the scanner must be tilted forward like in the picture bellow.



When you have tilted the scanner forward and find the correct position, please start from near to far until all bars on the right side are marked green. In the video bellow you can the the process.




Now step 4 need to be done. For this step the scanner must be tilted backward like in the picture bellow.



When you have tilted the scanner backward start from near to far until all bars on the right side are marked green. In the video bellow you can the the process.




Now step 5 need to be done. For this step the scanner must be tilted to the right like in the picture bellow.



When you have tilted the scanner to the right, start from near to far until all bars on the right side are marked green. In the video bellow you can the the process.





For the last step 6 the scanner must be tilted to the left like in the picture bellow.

 


When you have tilted the scanner to the left, start from near to far until all bars on the right side are marked green. In the video bellow you can the the process.





After all these steps has been done the calibration is finishes and the software start to calculate the result. If everthing was done correctly a new window should pop like bellow.




If the calibration fails please try again. If it fails the sec time please check with this article here, if you PC is set up correctly. When you PC is correctly set up but the calibration still fails please contact our support via metrology_support@shining3d.com